技術(shù)文章您的位置:網(wǎng)站首頁 >技術(shù)文章 >Easidew Pro I.S.露點(diǎn)儀 CE和UKCA認(rèn)證聲明

Easidew Pro I.S.露點(diǎn)儀 CE和UKCA認(rèn)證聲明

更新時(shí)間:2025-11-04   點(diǎn)擊次數(shù):37次

img1 

EU   Declaration  of  Conformity

 

 

Manufacturer:

Address:

 

 

 

 

img2Equipment Type:

 

 

Michell Instruments Limited

48 Lancaster Way Business Park Ely,Cambridgeshire

CB63NW.   UK.

img3 

 

Easidew   PRO   I.S.Dew-point   Transmitter

 

2014/34/EU   ATEX   Directive

Provisions of the Directive fulfilled by the Equipment:

Group  II  Category  1G       Ex      ia      IIC       T4-20℃≤Ta≤+70℃

Notified Body for Production(QAN &QAR):

SGS   Fimko   OY,Helsinki,Finland.Notified   Body   No.0598

Notified Body for EC-Type Examination:

SGS   Fimko   OY,Helsinki,Finland.Notified   Body   No.0598

EC-Type Examination Certificate:

Baseefa06ATEX0330X

Standards used:

EN60079-11:2012

EN       60079-0:2012/A11:2013

This standard has ceased to have harmonised status. EN60079-0:2018  has now superseded.A technical review of the latest version of this standard against the old standard showed that this equipment remains in conformance with allrelevant clauses and that the

State of the Art is maintained.The Essential Health &Safety Requirements of the Directive is sill maintained with no changes necessary for the safe and reliable functioning and operation of the product with respect to the risks of explosion.

 

IECEx

Certificate of Conformity No.

IECEx    BAS    06.0090X    Ex    ia    IIC    T4(-20≤Ta≤+70) IEC60079-0:2011

IEC60079-11:2011

Other  Directives   Is in conformity with the following Standard(s)or Normative Document(s):

2014/30/EU   EMC   Directive

EN61326-1:2013 Electrical equipment for measurement,control and laboratory use-EMC requirements.Class B(emissions)and Industrial Locations (immunity).

2011/65/EU-2015/863/EU               Restriction  of  Hazardous  Substances   Directive(RoHS3)

EU Directive 2015/863/EU,Annex II states:

The restriction of DEHP,BBP,DBP and DIBP shall apply to,……monitoring and control instruments,including industrial monitoring and control instruments,from 22 July 2021

EN61010-1:2010     Safety Requirements for Electrical Equipment for Measurement,Control,and Laboratory Use-Part 1:General Requirements 2014/68/EU   PE   Directive

This product and sample systems &accessories that may be supplied with it does not bear CE marking for the Pressure Equipment   Directive,and are supplied in accordance with Article 4,paragraph 3 of 2014/68/EU by using SEP(sound engineering practice)in the design and manufacturer and are provided with adequate instructions for use.

On behalf of the above named company,we the manufacturer declare under our sole responsibility that,on the date the equipment

accompanied by this declaration/s placed on the market,the equipment conforms with alltechnical and regulatory requirements of the above listed directives.

 

 

 

 

Andrew M.V.Stokes,Technical Director Febuary 2022

Ely,UK.                                                                                                                                                     ECD Easidew PRO IS Issue 13

 

UK    Declaration of Conformity

 

 

Manufacturer:

Address:

 

 

 

 

 

Equipment Type:

 

 

Michell Instruments Limited

48 Lancaster Way Business Park Ely,Cambridgeshire

CB63NW.UK.

 

 

 

Easidew    PROI.S.Dew-point    Transmitter

 

 

 

 

 

img4 

       image.png                                                                                  

The  Equipment  and  Protective Systems Intended for Use in Potentially Explosive Atmospheres Regulations     2016.UK     Statutory     Requirements(S.I.2016.1107.As     amended)

Provisions of the Directive fulfilled by the Equipment:

Group  II  Category  1G

Ex ia IIC T4-20℃≤Ta≤+70℃

Approved Body for UKCA-Type Examination

SGS   Baseefa.Buxton,UK    SK179RZ.Notified   Body    .1180

Approved Body for Production (UKCA QAN):

SGS   Baseefa.Buxton,UK    SK179RZ.Notified    Body   .1180

UKCA Type Examination Certificate:

BAS21UKEX0014X

Designated Standards used:

BS/EN60079-11:2012

BS/EN        60079-0:2012/A11:2013

This standard has ceased to have harmonised status. EN60079-0:2018    has now superseded.A technical review of the latest version of this standard against the old standard showed that this equipment remains in conformance with all relevant clauses and that the

State of the Art is maintained.The Essential Health &Safety Requirements of the Directive is still maintained with no changes necessary for the safe and reliable functioning and operation of the product with respect to the risks of explosion.

img5Other Regulations and Standards

The Electromagnetic Compatibility Regulations 2016.UK Statutory Instrument(S.I.2016 .1091).

Isin conformity with the following Standard(s)or Normative Document(s):

EN61326-1:2013 Electrical equipment for measurement,control and laboratory use-EMC requirements.Class B(emissions)and Industrial Locations (immunity).

UK Statutory Instrument(S.I.2012 No.3032).

The Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012.

States:

The  restriction  of  DEHP,BBP,DBP  and  DIBP  shall  apply  to,……monitoring  and  control  instruments,including  industrial  monitoring and  control  instruments,from  22  July  2021.

The Electrical Equipment Safety Regulation 2016.UK Statutory Instrument(S.I.20161101).BS/EN 61010-

1:2010/A1:2019       Safety Requirements for Electrical Equipment for Measurement,Control,and Laboratory Use -Part 1:General Requirements

The Pressure Equipment (Safety)Regulations 2016.UK Statutory Instrument (S.I.2016 .1105).

img6This product and sample systems &accessories that may be supplied with it does not bear UKCA marking for the Pressure Equipment(Safety) regulation,and are supplied in accordance with Article 4,paragraph 3 of 2014/68/EU by using SEP(sound engineering practice)in the design  and manufacturer and are provided with adequate instructions for use,

On behalf of the above named company,we  the  manufacturer  declare  under  our  sole  responsibility  that,on  the  date  the  equipment

accompanied  by this declaratiogn  is  placed  on the  market,the  equipment  conforms with all technical and  regulatory  requirements  of the

 

 

 

Andrew M.V.Stokes,Technical Director January 2022

Ely,UK.                                                                                                                        UKCA Easidew PRO IS Issue 02


聯(lián)